Kinetic™ Rhythms ECG Algorithm

The Kinetic™ Rhythms ECG Algorithm is designed to be embedded into a device or software suite to provide interpretation of up to sixteen distinct rhythms. It can be utilized in a variety of hardware platforms and applications due to its unique design and small footprint.

Kinetic™ Rhythms is specifically configured to provide precise interpretation of ECGs irrespective of the source of the data - whether taken at the bedside or in an ambulatory setting. It is able to discern true ECG from noise, leading to more accurate results - without all of the false positives.

The following table contains the test results of the Kinetic™ ECG Algorithm as compared to a gold standard. The database contains 250 ambulatory event monitor records that were sent trans-telephonically and then recorded. It has an equal dispersion of rhythms, and was interpreted by three cardiologists This data was used in our FDA 510K submission.

Rhythm
Sensitivity Positive Predictivity Specificity False Positive Rate
Normal Sinus Rhythm
91.18
100.00
100.00
0.00
Atrial Fibrillation
90.91
80.00
95.15
4.85
Nodal Rhythm
76.47
46.43
93.56
6.44
SVTA
100.00
61.90
92.86
7.14
Ventricular Tachycardia
100.00
85.71
99.59
0.41
BBB
82.14
79.31
97.30
2.73
1st Degree AV Block
88.46
74.19
96.43
3.57
2nd Degree AV Block (Mobitz 1)
96.77
100.00
100.00
0.00
2nd Degree AV Block (Mobitz 2)
100.00
100.00
100.00
0.00
Complete AV Block
96.77
100.00
100.00
0.00
Idioventricular Rhythm
72.73
88.89
99.58
0.42
Sinus Bradycardia
91.67
100.00
100.00
0.00
Sinus Tachycardia
77.78
100.00
100.00
0.00
APC
77.50
65.26
80.59
19.41
PVC
89.55
73.17
87.98
12.02
Pause
96.88
73.81
94.95
5.05
Overall
90.23
82.19
96.42
3.58